Published August 1998 | Version v1
Journal article

Plane polarized x-ray fluorescence system for the in vivo measurement of platinum in head and neck tumours

  • 1. Department of Medical Physics and Clinical Engineering, Singleton Hospital, Swansea SA2 8QA (United Kingdom)
  • 2. Department of Radiotherapy and Clinical Oncology, Singleton Hospital, Swansea SA2 8PP (United Kingdom)

Description

A plane polarized x-ray fluorescence system based on an orthovoltage radiotherapy treatment unit, the Pantak DXT-300, has been developed and optimized to measure tumour platinum concentration. The platinum derives from platinum based chemotherapy agents, such as cisplatin and carboplatin used to treat tumours in the head and neck region. Photons from an x-ray tube are polarized by scattering 90 deg., and used to stimulate the emission of characteristic platinum x-rays from the tumour. Information about the drug in the tumour could be of use in establishing dose-response relationships, in order to optimize the treatment and reduce the toxicity of the drug. The performance of the system was optimized with respect to the applied x-ray tube voltage, the current and the filter material; the effect on the minimum detection limit (MDL) of the thickness of the overlying tissue surrounding the tumour was investigated in detail. The lowest MDL is achieved using 0.25 mm of tin filter and x-ray tube voltage of 220 kV. This is 5.6 ppm for a tumour surrounded with 20 mm of overlying tissue, a measurement time of 2000 s and an estimated skin dose of 3.0 mGy. This represents the most sensitive in vivo XRF system to date. We have embarked on a clinical pilot study to measure the platinum concentration in tumours of the head and neck, and expect initial results to be available in the near future. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
43
Journal Issue
8
Journal Page Range
p. 2337-2345
ISSN
1361-6560

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43042001
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CARCINOMAS; CHEMOTHERAPY; DIAGNOSIS; HEAD; NECK; PLATINUM; RADIOTHERAPY; X-RAY FLUORESCENCE ANALYSIS
Descriptors DEC
BODY; CHEMICAL ANALYSIS; DISEASES; ELEMENTS; MEDICINE; METALS; NEOPLASMS; NONDESTRUCTIVE ANALYSIS; NUCLEAR MEDICINE; PLATINUM METALS; RADIOLOGY; THERAPY; TRANSITION ELEMENTS; X-RAY EMISSION ANALYSIS

Optional Information

Notes
23 refs; This record replaces 31036380