Published November 2014 | Version v1
Miscellaneous Open

Modified Liulin detector calibration for onboard aircraft dosimetry measurements

  • 1. Dept. of dosimetry, Nuclear Physics Institute of the ASCR, v. v. i., 25068 Rez (Czech Republic)
  • 2. Dept. of accelerators, Nuclear Physics Institute of the ASCR, v. v. i., 25068 Rez (Czech Republic)

Description

During their careers, aircraft crews are exposed to cosmic radiation more than people on the Earth's surface. Therefore, the exposure of aircraft crew to cosmic radiation is considered as occupational exposure in all EU member states, and even some non-EU countries. Routine dosimetry of aircraft crew is performed using computer programs based mainly on Monte Carlo simulations. Onboard dosimetry measurements are important for verification of the computational methods and for research. One of the detectors used for characterization of the radiation field onboard aircraft is active semiconductor detector Liulin originally developed for monitoring of cosmic radiation at the MIR space station. The output quantity of the measurement with the Liulin detector is the ambient dose equivalent H*(10) calculated with use of the absorbed dose D. This requires two calibrations. The purpose of our work was to develop one of these calibration methodologies converting the detector signal to the absorbed dose D independently of calibration performed at HIMAC facility in Chiba. The new calibration method is based on standard alpha radionuclide sources and replacement of the original PIN diode type S2744-08 with epoxy layer for the diode type S2744-09 without protective resin layer. The work reaches its conclusion by comparing the DSi results of Prague-Moscow flight, measured both by the Liulin detector with an original diode and standard calibration and the newly calibrated Liulin detector with the replaced diode. (authors)

Files

46116611.pdf

Files (3.0 MB)

Name Size Download all
md5:69b4d5fc078bb5798115a07f25ad3e12
3.0 MB Preview Download
Part of:
XXXVI. Days of Radiation Protection. Book of Presentations and Posters

Additional details

Additional titles

Original title (Czech)
Kalibrace upraveneho detektoru Liulin pro mereni kosmickeho zareni na palubach letadel

Publishing Information

Imprint Title
XXXVI. Days of Radiation Protection. Book of Presentations and Posters
Imprint Pagination
1171 p.
Journal Page Range
28 p.
Report number
INIS-SK--2015-003

Conference

Title
34. Days of Radiation Protection
Dates
10-14 Nov 2014
Place
Poprad (Slovakia)

Optional Information

Notes
27 figs., 1 ref. Imprint:XXXVI. Dni radiacnej ochrany. Kniha prezentacii a posterov