Comparison of four analytical techniques for determining uranium in different rock samples

  • 1. Nuclear Materials Authority, El Maadi, Cairo (Egypt)

Description

Four different analytical techniques were compared for their performance in direct determination of uranium in seven rock samples. These methods include: ion chromatography, laser fluorimetry, UV-Vis spectrophotometry and gamma spectrometry. The samples under study were six ore samples beside an international reference sample (GBWO4115). The results obtained from the comparison study reported a method accuracy of 98%, 90%, 65% and 86% for ion chromatography, laser fluorimetry, UV-Vis spectrophotometry and gamma spectrometry techniques respectively. Also, on the converse of the other three techniques, ion chromatography technique using DX-500 ion chromatograph achieves superior agreement with the certified sample with only 2% bias. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
322
Journal Issue
3
Journal Page Range
p. 1975-1982
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
36 refs.