Published August 1, 2016 | Version v1
Journal article

The influence of signals correlation on the long-term stability of a tandem of quantum magnetometers with laser pumping

  • 1. Institute of Physics, Nanotechnology and Telecommunications, Peter the Great Saint Petersburg Polytechnic University, St. Petersburg 195251 (Russian Federation)
  • 2. Ioffe Institute RAS, St. Petersburg 194021 (Russian Federation)

Description

The results of studies of the long-term frequency stability as a function of the correlation coefficient for a tandem of two quantum magnetometers with laser pumping of 87Rb in wall-coated vapour cell are represented. Measurement scheme includes a low-frequency self-generating magnetometer and a quantum microwave discriminator working at magnetic dipole transitions of radio-optical end state resonance. The difference of synchronously detected signals is processed to determine the Allan variance as a function of averaging time and correlation coefficient of signals. These parameters are essentially dependent both on the pumping light intensity and polarization and the intensity of radio fields that are produced in the working cell. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/741/1/012162

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
741
Journal Issue
1
Journal Page Range
[5 p.]
ISSN
1742-6596

Conference

Title
3. international school and conference on optoelectronics, photonics, engineering and nanostructures
Acronym
Saint Petersburg OPEN 2016
Dates
28-30 Mar 2016
Place