Published October 1992 | Version v1
Journal article

An automated thermal relaxation calorimeter for operation at low temperature (0.5K<T<10K)

  • 1. Indian Inst. of Science, Bangalore (India). Dept. of Physics

Description

An automated calorimeter for measurement of specific heat in the temperature range 10K>T>0.5K. It uses sample of moderate size (100-1000 mg), has a moderate precision and accuracy (2%-5%) is easy to operate and the measurements can be done quickly with 3He economy is described. The accuracy of this calorimeter was checked by measurement of specific heat of copper and that of aluminium near its superconducting transition temperature. (author). 12 refs., 11 figs

Additional details

Publishing Information

Journal Title
Pramana
Journal Volume
39
Journal Issue
4
Journal Page Range
p. 391-404.
ISSN
0304-4289
CODEN
PRAMCI