Published 1976 | Version v1
Report Open

Spectrographic determination of impurities in ammonium hydrogen fluoride samples

Description

The quantitative spectrographic trace determination of Al, B, Cr, Cu, Fe, Mn, Mo, Ni, Pb and Si in ammonium hydrogen fluoride samples is considered. 10 A dc arc excitation and graphite electrodes with crate either 4.5 mm or 8 mm deep are employed. A comparison of various matrices such as graphite, gallium oxide, germanium oxide, magnesium oxide and zinc oxide, in the ratios 1:1 and 1:3, as well as a mixture 50% graphite - 50% zinc oxide in the ratio 1:1 is included. Zinc oxide in the ratio 1:1 and 4x8 mm craters show the best over-all results. (author)

Availability note (English)

MF available from INIS under the Report Number.

Files

7243874.pdf

Files (142.9 kB)

Name Size Download all
md5:97e41fc1b5f70d31fdc71c6ec516ed9b
142.9 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Determinacion espectrografica de impurezas en bifluoruro amonico

Publishing Information

Imprint Pagination
11 p.
Report number
JEN--318

Optional Information

Notes
2 refs.