Accuracy of patient positioning in mantle field irradiation

  • 1. Dr. Daniel den Hoed Cancer Center, Rotterdam (Netherlands). Department of Radiation Oncology
  • 2. Dr. Daniel den Hoed Cancer Center, Rotterdam (Netherlands). Department of Clinical Physics

Description

A prospective study of the accuracy of patient positioning in mantle field irradiation was carried out in 13 lymphoma patients treated with curative radiotherapy. For anterior and posterior fields patients were treated in supine and prone position. Individually shaped divergent shielding blocks were placed in a fixed position in a template which was positioned on a tray above the patient. A total of 94 megavoltage portal films (MV) was analysed and compared to 26 simulation films (SIM). MV- SIM differences were larger for posterior than for anterior fields. Regarding the position of lung shielding blocks, mean MV-SIM differences ranged from 1.3-4.4 mm and errors exceeding 1 cm were found in 7.2% of the cases. Most discrepancies appeared to be randomly distributed. A 4-5 mm systematic cranial shift of patients in posterior treatment position was noted. Discrepancies in the position of the laryngeal block, spinal cord shielding block and humerus blocks were small with mean MV-SIM differences ranging from 0.3-2.7 mm. Differences between simulation set- up and treatment set-up were modest as compared to error rates reported in literature. Shielding of tumor-bearing areas did not occur. It was concluded that present standardised technique of patient positioning and design of treatment fields results in acceptable error rates. Attention should be directed towards increasing the stability of patients in the prone treatment position in order to further reduce both systematic and random error rates. (author). 13 refs.; 5 figs.; 3 tabs

Additional details

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
23
Journal Issue
4
Journal Page Range
p. 257-264.
ISSN
0167-8140
CODEN
RAONDT

INIS

Country of Publication
Ireland
Country of Input or Organization
Netherlands
INIS RN
24019248
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
ACCURACY; ERRORS; IRRADIATION; IRRADIATION DEVICES; LYMPHOMAS; NUMERICAL DATA; PHOTOGRAPHIC FILMS; POSITIONING; RADIOTHERAPY; SIMULATION
Descriptors DEC
DATA; DISEASES; IMMUNE SYSTEM DISEASES; INFORMATION; MEDICINE; NEOPLASMS; THERAPY