Published October 2004 | Version v1
Journal article

Enhanced sensitivity beam emission spectroscopy system for nonlinear turbulence measurements

  • 1. University of Wisconsin-Madison, 1500 Engineering Drive, Madison, Wisconsin 53706 (United States)

Description

An upgraded beam emission spectroscopy (BES) system has been deployed to access low amplitude turbulence regions near internal transport barriers on the DIII-D tokamak. Sixteen high sensitivity channels are being installed. A significant increase in total signal to noise is achieved by: (1) Increased spatial volume sampling tailored to known turbulence characteristics; (2) An increased throughput spectrometer assembly to isolate the local beam fluorescence, coupled to new large area photoconductive photodiodes; (3) A new sharp edge interference filter designed to optimize detection of the beam emission plus a significant fraction of the thermal deuterium charge exchange. A new data acquisition system has been installed, providing an eight times increase in integration time or an increased sample rate. Preliminary results from the upgraded system show a signal enhancement of greater than an order of magnitude. A clear broadband density fluctuation signal is observed in H-mode discharges with the upgraded BES system, demonstrating the significant performance enhancement

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
75
Journal Issue
10
Journal Page Range
p. 3493-3495
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2004 American Institute of Physics