Published May 21, 2016 | Version v1
Journal article

Operation of gas electron multiplier (GEM) with propane gas at low pressure and comparison with tissue-equivalent gas mixtures

  • 1. University of Padova, Physics and Astronomy Department and PD-INFN, via Marzolo 8, I-35131 Padova (Italy)
  • 2. Centre for Environmental Safety and Security, National Institute for Public Health and the Environment (RIVM), PO Box 1, NL-3720 BA Bilthoven (Netherlands)

Description

A Tissue-Equivalent Proportional Counter (TEPC), based on a single GEM foil of standard geometry, has been tested with pure propane gas at low pressure, in order to simulate a tissue site of about 1 µm equivalent size. In this work, the performance of GEM with propane gas at a pressure of 21 and 28 kPa will be presented. The effective gas gain was measured in various conditions using a 244Cm alpha source. The dependence of effective gain on the electric field strength along the GEM channel and in the drift and induction region was investigated. A maximum effective gain of about 5×103 has been reached. Results obtained in pure propane gas are compared with gas gain measurements in gas mixtures commonly employed in microdosimetry, that is propane and methane based Tissue-Equivalent gas mixtures.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2016.02.096

Additional details

Identifiers

DOI
10.1016/j.nima.2016.02.096;
PII
S0168-9002(16)30012-2;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
819
Journal Page Range
p. 154-162
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.