Lumbar spine radiography: lower organ dose with the use of PA projection
Creators
- 1. University of Ljubljana, Faculty of Health Sciences, Medical Imaging and Radiotherapy Department, Zdravstvena pot 5, Ljubljana (Slovenia)
Description
The purpose of the research was to determine the effect of the posteroanterior (PA) patient position in lumbar spine imaging on effective dose and the absorbed organ dose. The study was performed on 100 patients that were referred to the lumbar spine radiography that were divided into two equal groups of 50. Body Mass Index, Dose AREA Product (DAP), exposure index (EXI), tube time-current (mAs), image field size and the source-patient distance were acquired for each patient. The entrance surface dose (ESD), the effective dose and the absorbed organ doses were calculated. There was no statistically significant difference in the BMI and EXI between the AP and PA projection. The results showed a significant reduction of ESD by 33% and the effective dose by 53% when the PA projection was used. Furthermore, there was a 64% average reduction of the absorbed organ doses to the selected organs. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncz057Additional details
Identifiers
- DOI
- 10.1093/rpd/ncz057;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 186
- Journal Issue
- 4
- Journal Page Range
- p. 507-512
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 51102235
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DISTANCE; EFFECTIVE RADIATION DOSES; IMAGES; PATIENTS; VERTEBRAE
- Descriptors DEC
- BODY; DOSES; ORGANS; RADIATION DOSES; SKELETON
Optional Information
- Notes
- 31 refs.