Unicompartmental knee arthroplasty MRI: impact of slice-encoding for metal artefact correction MRI on image quality, findings and therapy decision
Creators
- 1. University of Zurich, Faculty of Medicine, Zurich (Switzerland)
- 2. Balgrist University Hospital, Radiology Department, Zurich (Switzerland)
- 3. Regional Hospital, Radiology, Lugano (Switzerland)
- 4. Balgrist University Hospital, Orthopedics, Zurich (Switzerland)
Description
To evaluate the impact of slice-encoding for metal artefact correction (SEMAC) on image quality, findings, and therapy decision in patients with unicompartmental knee arthroplasty (UKA). Forty-five painful UKAs were examined at 1.5T-MRI (STIR, proton-density(PD)-weighted sequence, each with SEMAC and high-bandwidth). Artefact size, image quality, anatomic depiction, and clinically relevant findings were compared between SEMAC and high-bandwidth (2 readers). In 30 patients, therapy decision was retrospectively assessed by two orthopaedic surgeons without MRI, with high-bandwidth-MRI, and with SEMAC-MRI. SEMAC reduced mean artefact size for STIR (11.8 cm2 vs. 37.7 cm2) and PD (16.8 cm2 vs. 18.9 cm2), p < 0.0005 for both comparisons. SEMAC showed more blurring than high-bandwidth, p < 0.0005. STIR-SEMAC revealed more bone marrow oedema (29 vs. 18 patients, p = 0.001, 30 vs. 13 patients, p < 0.0005, for reader 1 and 2 respectively). PD-SEMAC was worse in detecting meniscal lesions (6 missed, p = 0.031, 9 missed, p = 0.004, by reader 1 and 2 respectively) than PD-high-bandwidth. Revision-surgery was chosen in 12 and 11 patients without MRI (surgeon 1 and 2), with high-bandwidth-MRI in 15 and 14 patients, and with SEMAC-MRI in 19 and 14 patients. STIR-SEMAC was useful in detecting bone marrow oedema and influenced the orthopaedic surgeons' decisions towards surgery, while PD-SEMAC showed no clinical benefit. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-015-3596-4Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 25
- Journal Issue
- 7
- Journal Page Range
- p. 2184-2193
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 46105905
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BONE JOINTS; BONE MARROW; DECISION MAKING; EDEMA; IMAGE PROCESSING; IMPLANTS; LEGS; METALS; MUSCLES; NMR IMAGING; PAIN; PROTON DENSITY; SURGERY; THERAPY; WEIGHTING FUNCTIONS
- Descriptors DEC
- ANIMAL TISSUES; BODY; DIAGNOSTIC TECHNIQUES; ELEMENTS; FUNCTIONS; HEMATOPOIETIC SYSTEM; LIMBS; MEDICINE; ORGANS; PATHOLOGICAL CHANGES; PROCESSING; SKELETON; SYMPTOMS