Published April 2009 | Version v1
Journal article

Curved crystal x-ray optics for monochromatic imaging with a clinical source

  • 1. Department of Physics, University at Albany, SUNY, Albany, New York 12222 (United States)

Description

Monochromatic x-ray imaging has been shown to increase contrast and reduce dose relative to conventional broadband imaging. However, clinical sources with very narrow energy bandwidth tend to have limited intensity and field of view. In this study, focused fan beam monochromatic radiation was obtained using doubly curved monochromator crystals. While these optics have been in use for microanalysis at synchrotron facilities for some time, this work is the first investigation of the potential application of curved crystal optics to clinical sources for medical imaging. The optics could be used with a variety of clinical sources for monochromatic slot scan imaging. The intensity was assessed and the resolution of the focused beam was measured using a knife-edge technique. A simulation model was developed and comparisons to the measured resolution were performed to verify the accuracy of the simulation to predict resolution for different conventional sources. A simple geometrical calculation was also developed. The measured, simulated, and calculated resolutions agreed well. Adequate resolution and intensity for mammography were predicted for appropriate source/optic combinations.

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
36
Journal Issue
4
Journal Page Range
p. 1176-1183
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2009 American Association of Physicists in Medicine