Estimation of effective dose from CT examination
- 1. National Inst. of Radiological Sciences, Chiba (Japan)
Description
Tissue or organ doses related to radiological risk were determined for four different types of CT scanners with a spiral scan function. Dose measurements were performed using a Rando phantom and two types of thermoluminescent dosimeters. The effective doses recommended by the International Commission on Radiological Protection in 1990 were evaluated using the tissue or organ doses determined with the phantom measurement. The resultant effective dose per CT examination ranged from 4.6 to 10.8 mSv for chest examination and from 6.7 to 13.3 mSv for upper abdominal examination. It should be noted that the effective dose from CT examination will be increased by increasing in the frequency of CT examinations and technical development of CT scanners. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Igaku Hoshasen Gakkai Zasshi
- Journal Volume
- 55
- Journal Issue
- 11
- Journal Page Range
- p. 763-768.
- ISSN
- 0048-0428
- CODEN
- NHGZAR
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 27014034
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABDOMEN; BONE MARROW; CAT SCANNING; CHEST; COMPUTERS; DYNAMIC FUNCTION STUDIES; PATIENTS; PHANTOMS; RADIATION DOSES; SKIN; THERMOLUMINESCENT DOSIMETRY
- Descriptors DEC
- ANIMALS; BODY; BODY AREAS; COMPUTERIZED TOMOGRAPHY; DOSIMETRY; HEMATOPOIETIC SYSTEM; MAMMALS; MAN; MOCKUP; ORGANS; PRIMATES; STRUCTURAL MODELS; TISSUES; TOMOGRAPHY; VERTEBRATES