Published August 2012
| Version v1
Journal article
IMRT field profiling by high-quality polycrystalline CVD diamond
Creators
- 1. INFN Firenze and Dip. Energetica, University of Florence, Via S. Marta 3, Firenze (Italy)
- 2. INFN Firenze and Dip. Fisiopatologia Clinica, University of Florence, Largo Brambilla 3, Firenze (Italy)
- 3. INFN Firenze and Dip. Fisica e Astronomia, University of Florence, Sesto Fiorentino (Firenze) (Italy)
Description
A dosimeter made with state of art polycrystalline Chemical Vapour Deposited (pCVD) diamond for applications in clinical radiotherapy has been manufactured, based on a large area (2.5 × 2.5 cm2) film equipped with a matrix of contacts. A dose profile under an Intensity Modulated Radiotherapy (IMRT) beam for prostate cancer has been monitored using a set of in-line pads along a section of the field map. Results compare well with data obtained with a commercial bidimensional silicon dosimeter, confirming that our pCVD diamond device is suitable for mapping the conformed dose in IMRT treatments.
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/7/08/P08003Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 7
- Journal Issue
- 08
- Journal Page Range
- p. P08003
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44051148
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BEAMS; CHEMICAL VAPOR DEPOSITION; DIAMONDS; DOSEMETERS; EQUIPMENT; FILMS; NEOPLASMS; POLYCRYSTALS; PROSTATE; RADIATION DOSES; RADIOTHERAPY; SILICON
- Descriptors DEC
- BODY; CARBON; CHEMICAL COATING; CRYSTALS; DEPOSITION; DISEASES; DOSES; ELEMENTS; GLANDS; MALE GENITALS; MEASURING INSTRUMENTS; MEDICINE; MINERALS; NONMETALS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; SEMIMETALS; SURFACE COATING; THERAPY