Dosimetric study of 3D-printing coplanar template combined with CT-guided 125I seed implantation for treating non-small cell lung cancer

  • 1. Department of Oncology, The Second Hospital of Tianjin Medical University, Tianjin (China)
  • 2. Department of Thoracic Surgery, The Second Hospital of Tianjin Medical University, Tianjin (China)
  • 3. Department of Radiation Oncology, Peking University Third Hospital, Beijing (China)

Description

Objective: To explore the safety, effectivity and dosimetric continence of 3D-printing coplanar template(3D-PCT) combined with CT-guided 125I seed implantation in the treatment of non-small cell lung cancer(NSCLC). Methods: From May 2014 to November 2016, a total of 20 NSCLC patients who were suitable for 125I seed implantation were recruited in this study. Of all the patients, 10 received 125I seed implantation treatment by CT-guided combined with 3D-PCT (3D group), and the rest, by free-hand puncture combined with CT-guided 125I seed implantation (free-hand group). During two days before the surgery, the patients received the CT scan. Then the digital imaging and communications in medicine(DICOM) was collected to input to the Brachytherapy Treatment Planning System(BTPS). The dose parameters including D90, D100, V100, V150, conformal index(CI), external index(EI), and homogneneity index(HI) were compared between pre-operation and post-operation. Pair t-test and single sample t-test were performed. Results: V150 in 3D group between preoperation and postoperation showed statistically significant difference(t = -2.916, P < 0.05), and there was no significant difference in the rest parameters(P > 0.05). However, the number of seeds, V100, EI, HI in free-hand group between preoperation and postoperation showed statistically significant difference(t = -2.516, 2.492, 4.725, 7.258, P < 0.05), and there was no significant difference in the rest parameters(P > 0.05). Comparison of indicuted that there was significant difference in V100, V150, CI, EI, HI and dose error rate between the 3D group and the free-hand group with single sample t test, the result showed V100, CI, EI, HI between two groups(t = 2.598, 2.278, 4.637, 4.616, -4.785, P < 0.05), and there was no significant difference in V150 (P > 0.05). Conclusions: CT-guided 125I seed implantation brachytherapy combined with 3D-PCT for treatment of NSCLC safe is feasible, and dose controllable, and there is a certain advantage in the spatial distribution of seed compared with free-hand puncture. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Medicine and Protection
Journal Volume
37
Journal Issue
7
Journal Page Range
p. 528-532
ISSN
0254-5098

Optional Information

Notes
1 tab., 19 refs.; http://dx.doi.org/10.3760/cma.j.issn.0254-5098.2017.07.011