Published August 2012 | Version v1
Journal article

Evaluation of thermoluminescent BeO samples in standard radiotherapy beams

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

Description

Beryllium oxide thermoluminescent samples were evaluated in standard radiotherapy beams of low energy. Results for response reproducibility, dose-response curve and energy dependence were obtained. The lower detection limit was determined. The pellets of BeO showed their usefulness for beam dosimetry. (author)

Additional details

Additional titles

Original title (Portuguese)
Avaliacao de amostras termoluminescentes de BeO em feixes padroes de radioterapia

Publishing Information

Journal Title
Revista Brasileira de Fisica Medica (Online)
Journal Volume
6
Journal Issue
2
Journal Page Range
p. 95-98
ISSN
1984-9001