Book

In vivo measurement of actinides in the human lung

Description

The problems associated with the in vivo detection and measurement of actinides in the human lung are discussed with various measurement systems currently in use. In particular, the methods and calibration procedures employed at the Lawrence Livermore Laboratory, namely, the use of twin Phoswich detectors and a new, more realistic, tissue-equivalent phantom, are described. Methods for the measurement of chest-wall thickness, fat content, and normal human background counts are also discussed. Detection-efficiency values and minimum detectable activity estimates are given for three common actinides, 238Pu, 239Pu, and 241Am

Additional details

Publishing Information

Publisher
RD Press.
Imprint Place
Los Altos, CA
Imprint Title
Actinides in man and animals
Journal Page Range
p. 103-114.