Monitoring transverse beam profiles of a Penning ion source using a position-sensitive Multi Array Faraday Cup
Creators
- 1. Department of Radiation Application, Shahid Beheshti University G.C., Tehran (Iran, Islamic Republic of)
Description
In this paper, we describe the design and construction of a Multi Array Faraday Cup for both beam profiling and current measurements in a Penning ion source. This diagnostic system includes an array configured from multiple cups electrically isolated from each other to collect charged particles incident on the respective cups. Each cup in the array produces a specific signal corresponding to the collected charges due to the incident charged particles. Collected charges signals are amplified and processed by using an electronic beam monitoring system. A testing set-up has been designed for testing and calibration of the system. Experiments have been set up for evaluation of proper functioning of the MAFC in measurement of different beam profiles produced by using the testing set-up. Additionally, total extracted ion current and transverse beam profile has been measured for a Penning ion source
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/10/06/P06003Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 10
- Journal Issue
- 06
- Journal Page Range
- p. P06003
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47068410
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BEAM MONITORING; BEAM PROFILES; CALIBRATION; DESIGN; FARADAY CUPS; IONS; PENNING ION SOURCES; SIGNALS; TESTING
- Descriptors DEC
- BEAM MONITORS; CHARGED PARTICLES; ION SOURCES; MEASURING INSTRUMENTS; MONITORING; MONITORS