Absolute dose calibration of an X-ray system and dead time investigations of photon-counting techniques

Description

High precision concerning the dose calibration of X-ray sources is required when counting and integrating methods are compared. The dose calibration for a dental X-ray tube was executed with special dose calibration equipment (dosimeter) as function of exposure time and rate. Results were compared with a benchmark spectrum and agree within ±1.5%. Dead time investigations with the Medipix1 photon-counting chip (PCC) have been performed by rate variations. Two different types of dead time, paralysable and non-paralysable will be discussed. The dead time depends on settings of the front-end electronics and is a function of signal height, which might lead to systematic defects of systems. Dead time losses in excess of 30% have been found for the PCC at 200 kHz absorbed photons per pixel

Additional details

Identifiers

PII
S0168900202009488;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
487
Journal Issue
1-2
Journal Page Range
p. 71-77
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34008191
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
BENCHMARKS; CALIBRATION; COUNTING TECHNIQUES; DEAD TIME; PHOTONS; X-RAY DETECTION; X-RAY TUBES
Descriptors DEC
BOSONS; DETECTION; ELECTRON TUBES; ELEMENTARY PARTICLES; EQUIPMENT; MASSLESS PARTICLES; RADIATION DETECTION; TIMING PROPERTIES; X-RAY EQUIPMENT

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.