Three-dimensional dose distribution of tangential breast treatment
- 1. Dr. Daniel Den Hoed Cancer Centre, Rotterdam (Netherlands). Department of Clinical Physics
- 2. Nederlands Kanker Inst. 'Antoni van Leeuwenhoekhuis', Amsterdam (Netherlands)
Description
From Aug.'90-Feb.'91, dosimetry intercomparison of breast treatment was performed at all 21 radiotherapy centres in The Netherlands. The absorbed dose was measured in three planes in a breast phantom during tangential breast irradiation, according to a prescribed technique. The beam energy could be chosen by the radiotherapy centre as normally applied for this type of 'patient', and varied between 60Co and 8MV X-rays. The dose measured by the visiting team in 22 points inside the phantom was compared with the dose calculated by the institution using their local treatment planning system. In the institutions mean ratio (mean value of ratios of the absolute calculated dose and measured absolute dose in the 22 points) varied between 0.92-1.08 with an over-all mean ratio of 1.04. There was no significant difference in this ratio between the 3 planes in a particular institution. In the isocentre mean ratio of calculated and measured dose was 1.021 with a SD of 0.028 i.e. the algorithms in the 6 different commercial treatment planning systems calculate the dose generally somewhat too high. In order to explain results, a measurement of the output under reference conditions was performed at each treatment unit. Mean ratio of the dose stated by the institution and the dose measured by the visiting team was 1.011 with a 0.015-SD with a maximum deviation of 0.040. This small deviation explains therefore only part of variation in the ratio of calculated and measured dose for tangential breast irradiation. In several centres large deviations between actual beam data and beam data implemented in the planning system were found, which was the main reason for observed discrepancies. (author). 15 refs.; 5 figs.; 2 tabs
Additional details
Additional titles
- Subtitle (English)
- A national dosimetry intercomparison
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 22
- Journal Issue
- 4
- Series
- Radiother. Oncol.
- Journal Page Range
- 252-260
- ISSN
- 0167-8140
- CODEN
- RAOND
Conference
- Title
- A Consensus Meeting on Quality Assurance in Conservative Treatment of Early Breast Cancer.
- Dates
- 13-14 Dec 1990.
- Place
- Tuebingen (Germany).
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- Netherlands
- INIS RN
- 23058730
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DOSIMETRY; MAMMARY GLANDS; NEOPLASMS; SPATIAL DOSE DISTRIBUTIONS
- Descriptors DEC
- BODY; DISEASES; DISTRIBUTION; GLANDS; ORGANS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION
Optional Information
- Contract/Grant/Project number
- Grant NKI 89-10
- Notes
- This work was supported by NKB grant from the Netherlands Cancer Foundation.