Published March 2018 | Version v1
Journal article

FLUKA-based Monte Carlo investigation of Microdosimetric distributions of telecobalt beam

  • 1. Radiological Physics and Advisory Division, Health, Safety and Environment Group, Bhabha Atomic Research Centre, Mumbai 400094 (India)

Description

FLUKA-based Monte Carlo calculations of microdosimetric distributions in water phantom involving a walled spherical Tissue-Equivalent Proportional Counter filled with tissue-equivalent propane gas have been studied for an indigenously developed telecobalt machine. The simulated site size considered in the study was 2 μm. In the Monte Carlo calculations, field size was varied from 10 cm x 10 cm to 35 cm x 35 cm and the depth was varied as 5-20 cm. The study also includes calculation of microdosimetric distributions with a 30 deg. wedge filter. The efficiency of the calculations was improved up to a factor of 26 by choosing appropriate cut off values for production and transport of electrons. The calculated microdosimetric distributions of telecobalt machine is distinctly different from that of a bare 60Co source which is attributed to the influence of scattered photons from the machine head and the water phantom. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncx182

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
178
Journal Issue
4
Journal Page Range
p. 430-440
ISSN
0144-8420

Optional Information

Notes
30 refs.