Physical characteristics of the Selectron high dose rate intracavitary afterloader
- 1. McGill Univ., Montreal, Quebec (Canada)
- 2. Royal Victoria Hospital, Montreal, Quebec (Canada)
Description
The physics measurements on a Selectron high dose-rate afterloading cobalt-60 unit are reported. The installation was found to be acceptable from the standpoint of radiation safety and cost effectiveness; hospital bed space was saved as treatment could be on an outpatient basis. A source calibration 4% higher than the value stated by the manufacturer was obtained. Measurement of the ratio of exposure rate in water to that in air confirmed the calibration and the applicability of correction factors for routine clinical dosimetry recommended in the literature. (author)
Additional details
Publishing Information
- Journal Title
- Br. J. Radiol.
- Journal Volume
- 58
- Journal Issue
- 692
- Series
- Br. J. Radiol.
- Journal Page Range
- 735-740
- ISSN
- 0007-1285
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16077861
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AFTERLOADING; CALIBRATION; COBALT 60; CORRECTIONS; COST; DOSE RATES; DOSIMETRY; INTERNAL IRRADIATION; PERFORMANCE TESTING; RADIATION PROTECTION; RISK ASSESSMENT
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; IRRADIATION; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; RADIOTHERAPY; TESTING; THERAPY; YEARS LIVING RADIOISOTOPES