Performance of mammography equipment in the macedonian breast screening campaign 2008/2009
Creators
- 1. Inst. of Physics, Faculty of Natural Science and Mathematics, Gazi Baba bb, 1000 Skopje (Macedonia, The Former Yugoslav Republic of)
- 2. Univ. Clinic of Radiology, Vodnjanska 17, 1000 Skopje (Macedonia, The Former Yugoslav Republic of)
Description
Breast screening campaign in Macedonia started in the end of 2007 and 19 national mammography departments were included. Contrary to the European Guidelines for Quality Assurance in Mammography Screening, the quality assurance activities were not implemented before the start of the campaign, except at the Univ. Clinic of Radiology, Skopje. The quality control tests were performed for the very first time at 13 mammography units under a licence-obtaining procedure. One of the machines was suspended from clinical and screening practice due to heavy malfunction of the generator, X-ray tube and automatic exposure control (AEC) system. Only 3 of the 13 mammography machines met the criteria for tube voltage (kV) accuracy. Two of the seven AEC systems were calibrated in the optimal optical density (OD) range (OD > 1.4). AEC settings corresponded to the recommendations at eight units, while nine units met basic overall image quality criteria. Mean glandular dose (MGD) was higher than the recommended level of 2.5 mGy in four departments. Mean gradient of the film G 0.25-2.0 was below 2.8 at four units. Only two light boxes had a luminance of > 1700 cd m-2 and six rooms had an ambient light level of > 50 lx. The findings of this work clearly suggest that the performance of the mammography equipment involved in the campaign in almost 50 % do not supply basic quality criteria for a breast screening programme. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncr290Additional details
Identifiers
- DOI
- 10.1093/rpd/ncr290;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 147
- Journal Issue
- 1-2
- Journal Page Range
- p. 187-191
- ISSN
- 0144-8420
Conference
- Title
- International Conference on Radiation Protection in Medicine
- Dates
- 1-3 Sep 2010
- Place
- Varna (Bulgaria)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43077820
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; BIOMEDICAL RADIOGRAPHY; DOSES; FEMALES; HUMAN POPULATIONS; MAMMARY GLANDS; NEOPLASMS; QUALITY ASSURANCE; QUALITY CONTROL; RECOMMENDATIONS; SCREENING; X-RAY TUBES
- Descriptors DEC
- BODY; CONTROL; DIAGNOSTIC TECHNIQUES; DISEASES; ELECTRON TUBES; EQUIPMENT; GLANDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; POPULATIONS; RADIOLOGY; X-RAY EQUIPMENT
Optional Information
- Notes
- 3 refs