Published 1983 | Version v1
Report

Response of a high-pressure ionization chamber in dependence on filling gas and radiation quality

Description

Studies on the characteristics of ionization chambers as a function of the pressure of various filling gases are of interest in radiation protection. The tissue equivalent (TE) ionization chambers used for dosimetry in radiobiology have too low a sensitivity for application in radiation protection and provide no information on radiation quality. These restrictions might be resolved by operating ionization chambers at higher gas pressures. With the aim of obtaining a detector with a greater sensitivity, the reading of the high-pressure ionization chamber has been studied for other gases by use of 137Cs γ rays. For various gases, the relative reading as a function of gas pressure is shown. (Auth.)

Additional details

Publishing Information

Imprint Title
REP annual report 1983
Imprint Pagination
275 p.
Journal Page Range
p. 3-5.
Report number
TNO-REP--1983

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
17013520
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Progress Report
Descriptors DEI
GASES; IONIZATION CHAMBERS; PERFORMANCE TESTING; RADIATION PROTECTION; SENSITIVITY
Descriptors DEC
FLUIDS; MEASURING INSTRUMENTS; RADIATION DETECTORS; TESTING

Optional Information

Notes
3 refs.; 2 figs.