Temperature calibration of Pico-Rad detectors for radon measurement

  • 1. Institute of Applied Radiation Chemistry, Lodz, (Poland)
  • 2. Department of Environmental Engineering Technical University, Lodz (Poland)
  • 3. Department of Radiation Protection the Nofer of Occupational Medicine, Lodz (Poland)

Description

A simple mathematical equation linking the activity of adsorbed radon in the vials to the time and temperature of its exposure is discussed. The calibration coefficient - Ks, defined as activity measured in cpm after saturation time, corresponding to radon air concentration of 1 Bq m-3, was determined for four temperatures: 284, 291, 294 and 298 K. A linear relationship of ln Ks values versus T-1 was found. The relatively high difference in Ks values: 2.12 and 1.24 cpm/Bq m-3 for the temperatures of 284 and 298 K, respectively, was observed. It indicates that temperature fluctuations during Pico-Rad vial exposure may lead to erroneous results if the constant average temperature of exposure is introduced into a commonly used computer programme for calculating Rn concentration. (author)

Additional details

Publishing Information

Journal Title
International Journal of Occupational Medicine and Environmental Health
Journal Volume
13
Journal Issue
2
Journal Page Range
p. 147-154
ISSN
1232-1087

Optional Information

Notes
9 refs, 5 figs