Temperature calibration of Pico-Rad detectors for radon measurement
- 1. Institute of Applied Radiation Chemistry, Lodz, (Poland)
- 2. Department of Environmental Engineering Technical University, Lodz (Poland)
- 3. Department of Radiation Protection the Nofer of Occupational Medicine, Lodz (Poland)
Description
A simple mathematical equation linking the activity of adsorbed radon in the vials to the time and temperature of its exposure is discussed. The calibration coefficient - Ks, defined as activity measured in cpm after saturation time, corresponding to radon air concentration of 1 Bq m-3, was determined for four temperatures: 284, 291, 294 and 298 K. A linear relationship of ln Ks values versus T-1 was found. The relatively high difference in Ks values: 2.12 and 1.24 cpm/Bq m-3 for the temperatures of 284 and 298 K, respectively, was observed. It indicates that temperature fluctuations during Pico-Rad vial exposure may lead to erroneous results if the constant average temperature of exposure is introduced into a commonly used computer programme for calculating Rn concentration. (author)
Additional details
Publishing Information
- Journal Title
- International Journal of Occupational Medicine and Environmental Health
- Journal Volume
- 13
- Journal Issue
- 2
- Journal Page Range
- p. 147-154
- ISSN
- 1232-1087
INIS
- Country of Publication
- Poland
- Country of Input or Organization
- Poland
- INIS RN
- 36110498
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- AIR; CALIBRATION STANDARDS; COMPUTER CODES; ENERGY SPECTRA; LIQUID SCINTILLATION DETECTORS; MATHEMATICAL MODELS; RADIATION DOSES; RADIATION MONITORING; RADON 222; TEMPERATURE MONITORING
- Descriptors DEC
- ALPHA DECAY RADIOISOTOPES; DAYS LIVING RADIOISOTOPES; DOSES; EVEN-EVEN NUCLEI; FLUIDS; GASES; HEAVY NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; MONITORING; NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; RADON ISOTOPES; SCINTILLATION COUNTERS; SPECTRA; STANDARDS
Optional Information
- Notes
- 9 refs, 5 figs