Published June 1982
| Version v1
Journal article
Evaluation of the weighted geometric center in the intracavitary therapy for carcinoma of the uterine cervix using remote afterloader
- 1. Osaka Prefectural Center for Adult Diseases (Japan)
Description
In the intracavitary technique, the border of the target volume is generally determined by the local anatomy and applicators employed. Since the non-rigid applicator is used at our department, there are some differences between point A dose on both sides, and also between the planned and delivered doses of point A. Based on this study, it is suggested that the weighted geometric center is reflected in the inclination of the dose distribution and the change of mgh. It seems likely that the weighted geometric center is one of parameters to compare the intracavitary technique for carcinoma of the uterine cervix. (author)
Additional details
Publishing Information
- Journal Title
- Rinsho Hoshasen
- Journal Volume
- 27
- Journal Issue
- 6
- Series
- Rinsho Hoshasen.
- Journal Page Range
- 675-679
- ISSN
- 0009-9252
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 14779618
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AFTERLOADING; CARCINOMAS; IRRADIATION CAPSULES; IRRADIATION PROCEDURES; PATIENTS; RADIATION SOURCE IMPLANTS; SPATIAL DOSE DISTRIBUTIONS; UROGENITAL SYSTEM DISEASES; UTERUS
- Descriptors DEC