Published August 2019 | Version v1
Miscellaneous Open

Characteristics of simulated workplace neutron standard fields

  • 1. Institute for Nuclear Science and Technology, VINATOM, 179 Hoang Quoc Viet, Hanoi (Viet Nam)

Description

This paper presents the development of simulated workplace neutron standard fields at the Institute for Nuclear Science and Technology with the 241Am-Be source moderated by polyethylene spheres with diameters of 15 cm and 30 cm. The characterization of the standard fields (in terms of neutron fluence rates and neutron ambient dose equivalent rates) was performed using Bonner sphere spectrometer system together with MAXED and FRUIT unfolding codes. The related quantities such as neutron dose equivalent-averaged energies and fluence-to-ambient dose equivalent conversion coefficients were also determined. The discrepancies of values are satisfied the standard uncertainty criteria as recommended by the International Standard Organization 12789 series. It implies that the simulated workplace neutron standard fields can be applied in the practical works for calibration purposes. (author)

Files

51081736.pdf

Files (73.7 MB)

Name Size Download all
md5:619a5762db0cd256f507f70bb20097fa
73.7 MB Preview Download
Part of:
Vietnam Conference on Nuclear Science and Technology VINANST-13. Agenda and Abstracts

Additional details

Publishing Information

Imprint Title
Vietnam Conference on Nuclear Science and Technology VINANST-13. Agenda and Abstracts
Imprint Pagination
266 p.
Journal Page Range
8 p.
Report number
INIS-VN--004

Conference

Title
13. Vietnam Conference on Nuclear Science and Technology
Original Conference Title
Hoi nghi Khoa hoc va Cong nghe Hat nhan Toan quoc lan thu 13
Acronym
VINANST-13
Dates
7-9 Aug 2019
Place
Ha Long City, Quang Ninh (Viet Nam)

INIS

Country of Publication
Viet Nam
Country of Input or Organization
Viet Nam
INIS RN
51081736
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Resource subtype / Literary indicator
Conference
Descriptors DEI
AMBIENT DOSE EQUIVALENTS; BONNER SPHERE SPECTROMETERS; CALIBRATION; CONVERSION; NEUTRON FLUENCE; UNCERTAINTY PRINCIPLE
Descriptors DEC
DOSE EQUIVALENTS; MEASURING INSTRUMENTS; NEUTRON SPECTROMETERS; SPECTROMETERS

Optional Information

Notes
11 refs, 6 figs; This record replaces 51052853 Imprint:Conference program and paper abstracts only; Section C: Radiation Measurement, Radiation Safety And Environmental Monitoring