Contaminant electrons in the build-up region of a 4 MV photon beam
- 1. George Washington Univ. Medical Center, Washington, DC (USA). Dept. of Radiology
- 2. Peninsula General Hospital Medical Center, Salisbury, MD (USA). Dept. of Radiation Oncology
Description
The per cent ionization curves for the 4 MV photon beam from a Clinac 4 exhibit pronounced enhancement with field size in the build-up region. The source and nature of this increase is investigated by analysing the radiation components of the 4 MV photon beam with a 0.23 T magnetic field. The build-up ionization curves were measured at 86 cm SSD with a parallel-plate ionization chamber embedded in a polystyrene phantom. The surface ionization increases from 18% to 32% as the field size is increased from 4.3 cm x 4.3 cm to 9.7 cm x 32.3 cm. With the electromagnet placed beside the treatment head, the field size dependence reduces to 16% to 24%. This and similar results at 113.2 cm SSD, clearly demonstrate that electrons from the Clinac 4 treatment head are responsible for approximately half the observed increase in the surface ionization with field size. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Issue
- 28(6
- Series
- Phys. Med. Biol.
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 14806264
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE; S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- CONTAMINATION; ELECTRONS; IONIZATION; IONIZATION CHAMBERS; MEV RANGE 01-10; NONUNIFORM IRRADIATION; PHANTOMS; PHOTON BEAMS; RADIOTHERAPY; SPATIAL DOSE DISTRIBUTIONS
- Descriptors DEC
- BEAMS; DISTRIBUTION; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; IRRADIATION; LEPTONS; MEASURING INSTRUMENTS; MEDICINE; MEV RANGE; MOCKUP; RADIATION DETECTORS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; STRUCTURAL MODELS; THERAPY