Published June 1983 | Version v1
Journal article

Contaminant electrons in the build-up region of a 4 MV photon beam

  • 1. George Washington Univ. Medical Center, Washington, DC (USA). Dept. of Radiology
  • 2. Peninsula General Hospital Medical Center, Salisbury, MD (USA). Dept. of Radiation Oncology

Description

The per cent ionization curves for the 4 MV photon beam from a Clinac 4 exhibit pronounced enhancement with field size in the build-up region. The source and nature of this increase is investigated by analysing the radiation components of the 4 MV photon beam with a 0.23 T magnetic field. The build-up ionization curves were measured at 86 cm SSD with a parallel-plate ionization chamber embedded in a polystyrene phantom. The surface ionization increases from 18% to 32% as the field size is increased from 4.3 cm x 4.3 cm to 9.7 cm x 32.3 cm. With the electromagnet placed beside the treatment head, the field size dependence reduces to 16% to 24%. This and similar results at 113.2 cm SSD, clearly demonstrate that electrons from the Clinac 4 treatment head are responsible for approximately half the observed increase in the surface ionization with field size. (author)

Additional details

Publishing Information

Journal Title
Phys. Med. Biol.
Journal Issue
28(6
Series
Phys. Med. Biol.
ISSN
0031-9155