By using feedback control of the discrimination threshold we have added 4-6 decades to the useful range of otherwise conventional pulse counting systems. The additional range is obtained at the high end by retarding saturation of the output count rate with higher threshold levels. The authors have implemented the method with low-power microcircuits that provide a 9-decade range (10/sup -5/ to 10-/sup 4/ R/h) without the need for range or mode switching
Additional details
Publishing Information
- Journal Title
- IEEE Trans. Nucl. Sci.
- Journal Volume
- NS-34
- Journal Issue
- 1
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 78-81
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- Nuclear science and nuclear power systems symposium.
- Dates
- 29-31 Oct 1986.
- Place
- Washington, DC (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18066166
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S42: ENGINEERING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CONTROL SYSTEMS; COUNTING RATEMETERS; COUNTING RATES; FEEDBACK; MICROELECTRONIC CIRCUITS; PERFORMANCE; PULSE PILEUP; PULSE TECHNIQUES; RADIATION DETECTORS; SATURATION
- Descriptors DEC
- ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS
Optional Information
- Secondary number(s)
- CONF-861007--.