Feedback threshold control extends pulse rate measurements into the pileup region

  • 1. Instrumentation and Controls Div., Oak Ridge National Lab., Bldg. 3500, Oak Ridge, TN 37831-6008

Description

By using feedback control of the discrimination threshold we have added 4-6 decades to the useful range of otherwise conventional pulse counting systems. The additional range is obtained at the high end by retarding saturation of the output count rate with higher threshold levels. The authors have implemented the method with low-power microcircuits that provide a 9-decade range (10/sup -5/ to 10-/sup 4/ R/h) without the need for range or mode switching

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-34
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
78-81
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Nuclear science and nuclear power systems symposium.
Dates
29-31 Oct 1986.
Place
Washington, DC (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
18066166
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S42: ENGINEERING;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CONTROL SYSTEMS; COUNTING RATEMETERS; COUNTING RATES; FEEDBACK; MICROELECTRONIC CIRCUITS; PERFORMANCE; PULSE PILEUP; PULSE TECHNIQUES; RADIATION DETECTORS; SATURATION
Descriptors DEC
ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS

Optional Information

Secondary number(s)
CONF-861007--.