Published May 2000 | Version v1
Miscellaneous

Development of position sensitive neutron detector and study of its characteristics

  • 1. Kyungpook National Univ., Daegu (Korea, Republic of)
  • 2. KAREI, Taejon (KR)

Description

A one dimensional position sensitive neutron detector was designed and fabricated, which can reduce measurement time drastically when the scattering power of sample is very low. The incident neutron position was encoded by delay lines which are connected to each cathode strips. The effective detection area of the detector is 120mm x 80mm and the thermal neutron detection efficiency is about 60%. The intrinsic spatial resolution is 1.6mm. We could get 1.9mm of spatial resolution using a 1mm width slit. The integral non-linearity of the detector and the differential non-linearity of the detector are about 0.3% and 3.67%, respectively. The fabricated detector was applied to the diffraction measurement to see feasibility of field application, and the acquired Ni powder diffraction patterns were compared with the data from the HRPD(High Resolution Powder Diffractometer) in the research reactor HANARO

Part of:
Proceedings of the KNS spring meeting

Additional details

Publishing Information

Publisher
KNS
Imprint Place
Taejon (Korea, Republic of)
Imprint Title
Proceedings of the KNS spring meeting
Imprint Pagination
[CDROM]
Journal Page Range
[14 p.]

Conference

Title
2000 spring meeting of the KNS
Dates
26-27 May 2000
Place
Kori (Korea, Republic of)

Optional Information

Notes
6 refs, 15 figs, 1 tab