Published May 2009 | Version v1
Journal article

Quantification of methotrexate by liquid chromatography ultraviolet detection for routine monitoring of plasma levels

  • 1. Centre National de Pharmacovigilance (Tunisia)

Description

A high-performance liquid chromatographic (HPLC) technique with ultraviolet detection incorporating solid phase extraction (SPE) was developed to meet analytical and metrological requirements for routine serum level monitoring of methotrexate (MTX), with several parameters optimised such as temperature, flow rate, composition of the mobile phase and pH of the buffer solution. Two standard curves were constructed to cover the high and low levels of the calibrator range (0.02-600 micro mol/litre). Reproducibility (precision) of the method for intra assay was 2.7; 2.10; 1.38% at the lowest level and 2.11; 3.4; 2.01% at the highest level and for inter assay was 2.8; 2.2; 2.94% at the lowest level and 2.4; 2.74; 2.68% at the highest level; recovery was between 90.47 and 98.53 percent. Response was found linear over the whole range of the calibrator set with a correlation coefficient of 0.999. The limit of quantification and the limit of detection were 0.02 micro mol/litre and 0.0063 micro mol/litre, respectively. The method is suitable for quantification of methotrexate with good accuracy and precision. (author)

Additional details

Publishing Information

Journal Title
Pakistan Journal of Scientific and Industrial Research
Journal Volume
52
Journal Issue
3
Journal Page Range
p. 154-157
ISSN
0030-9885