Simple home-made sensors for potentiometric titrations

Description

A sensor for potentiometric titrations was prepared by coating a spectroscopic graphite rod with a solution of poly(vinyl chloride) and dioctyphthalate in tetrahydrofuran. The reference electrode was an Ag/AgCl single-junction electrode. The sensor was used in precipitation, acid-base, compleximetric, and redox titrations. Preparation of the coated-graphite sensor is simple and rapid. Moreover, it is quite inexpensive. A limitation is its applicability in aqueous media only, because organic solvents will dissolve the membrane. Various uncoated types of graphite have also been investigated as sensors, particularly in two applications of interest in the analysis of propellants: the titration of nitroform and perchlorate. Obviously, these sensors can be used also in nonaqueous, or partially nonaqueous media. These sensors may also find use in the potentiometric titration of fluoride vs La(III) or Th

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01 as DE82013792.