Published May 2002 | Version v1
Journal article

SQUID-based measuring systems

  • 1. Materials Science Div., Indira Gandhi Centre for Atomic Research, Kalpakkam (India)

Description

A program has been developed and initiated at the Indira Gandhi Centre for Atomic Research (IGCAR) for the utilization of SQUID sensors in various application areas. DC SQUID sensors based on Nb-AlOx-Nb Josephson junctions have been designed and developed in house along with associated flux-locked loop (FLL) electronics. A compact low field SQUID magnetometer insertible in a liquid helium storage dewar has also been developed in house and is in use. Efforts to build a high field SQUID magnetometer, SQUID-DAC system, are in progress. A planar gradiometric DC SQUID sensor for non-destructive evaluation (NDE) application to be used in relatively unshielded environment has been designed and developed. An easily portable NDE cryostat with a small lift-off distance, to be used in external locations has been designed and tested. The magnetic field produced by a given two-dimensional current density distribution is inverted using the Fourier transform technique. (author)

Additional details

Publishing Information

Journal Title
Pramana
Journal Volume
58
Journal Issue
5-6
Journal Page Range
p. 1159-1164
CODEN
PRAMCI

Conference

Title
advances in superconductivity and magnetism: materials, mechanisms and devices
Acronym
ASMM2D-2001
Dates
25-28 Sep 2001
Place
Mangalore (India)

Optional Information

Notes
6 refs., 2 figs.