Published November 1976 | Version v1
Journal article

Neutron activation analysis program for multielement trace analysis with absolute counting

  • 1. Bell Telephone Labs., Inc., Allentown, PA

Description

A Neutron Activation Analysis (NAA) program using absolute counting and a new method of determination of the thermal and epithermal flux has been developed. Only two standards (Fe and Ru) are needed, and only the weight of the Fe must be known accurately. The program runs in a computer with 16K memory. The program is primarily designed for the nondestructive quantitative analysis of impurities at the μg/g--ng/g level in silicon or other high purity matrices which do not themselves remain very radioactive long after end of irradiation. Otherwise removal of dominant activities prior to counting is necessary. Accuracy for isotopes with well-established nuclear constants is better than 10%. All features of the program are designed to enable the user to co-irradiate and evaluate a large number of samples per activation run in order to keep the irradiation and analysis cost per sample as low as possible

Additional details

Identifiers

Publishing Information

Journal Title
Analytical Chemistry
Journal Volume
48
Journal Issue
13
Series
Anal. Chem.
Journal Page Range
1962-1969
ISSN
0003-2700

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent