Numerical simulation of photon response in portal monitors
Creators
- 1. Sandia National Laboratories, Project Engineering Division, Albuquerque, New Mexico 87185
Description
The adjoint technique has been applied to accurately and economically predict the response of a portal monitor to photon emissions below about 1.5 MeV, thus encompassing those sources generally of interest in nuclear safeguards applications. The adjoint source was defined as the product of the total attenuation coefficient and an experimentally determined efficiency factor, which accounts for the performance characteristics of the signal-processing system. The efficiency factor was determined from a combination of data obtained from a single NE-102 scintillator and results from corresponding three-dimensional forward MORSE calculations. A prototype walk-through portal was then fabricated with four identical NE-102 scintillators. Adjoint MORSE calculations were performed to obtain net count rates for various sources within this portal. These results were compared to experimental data and were found to agree to well within 10%. The photon response within the portal detection volume was then characterized by a series of MORSE calculations
Additional details
Publishing Information
- Journal Title
- Nucl. Technol.
- Journal Volume
- 67
- Journal Issue
- 2
- Series
- Nucl. Technol.
- Journal Page Range
- 333-340
- ISSN
- 0029-5450
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 16061987
- Subject category
- S98: NUCLEAR DISARMAMENT, SAFEGUARDS AND PHYSICAL PROTECTION; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPUTERIZED SIMULATION; DATA PROCESSING; PERFORMANCE; RADIATION MONITORS
- Descriptors DEC
- MEASURING INSTRUMENTS; MONITORS; SIMULATION