Published March 2013 | Version v1
Journal article

Energy calibration for LaBr3(Ce) scintillator detector in the region of 1-10 MeV

  • 1. Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang (China)
  • 2. College of Physical Science and Technology, Key Laboratory for Radiation Physics and Technology of Ministry of Education, Sichuan University, Chengdu (China)

Description

Background: LaBr3(Ce) detector has played an important role in detecting explosive, contraband and landmine because of its high y detection efficiency and good energy resolution etc. Purpose: To calibrate detector in wide energy region. Methods: The gamma spectra of NH4Cl and C3H6N6 induced by 252Cf neutron source were measured. Results: Comparing their gamma spectra, characteristic gamma lines can be located and the energy calibration curve was obtained. Conclusions: Radio nuclides can be identified by the calibration curve fitted with quadratic or cubic polynomial. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Techniques
Journal Volume
36
Journal Issue
3
Journal Page Range
[4 p.]
ISSN
0253-3219

Optional Information

Notes
3 figs., 4 tabs., 6 refs., 030204-1-030204-4