Preparation and calibration of W5%Re v. W26%Re thermocouples for irradiation in H. F. R. Petten

  • 1. UKAEA Research Group, Harwell. Atomic Energy Research Establishment

Description

The calibration of W Re thermocouples under neutron irradiation has been studied at Petten in a collaborative experiment. The experiment was designed to separate the effects of transmutation from those of displacement damage by running at temperatures (1500 deg C) high enough for the displacement damage to anneal out. The methods and conditions of manufacture of some of the thermocouples and also the calibration techniques used are described. Travelling gradient calibration was used to identify the position of inhomogeneities in the wires before irradiation and will be compared with similar data after irradiation

Additional details

Publishing Information

Imprint Title
International colloquium on high-temperature in-pile thermometry, held at J.R.C. Petten, Netherlands, 12-13th December 1974
Imprint Pagination
p. 363-375.
Report number
EUR--5395(vol.1)

Conference

Title
International colloquium on high-temperature in-pile thermometry.
Dates
12 Dec 1974.
Place
Petten, Netherlands.

INIS

Country of Publication
European Commission (EC), Brussels (Belgium)
Country of Input or Organization
European Commission (EC), Brussels (Belgium)
INIS RN
7249486
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; FABRICATION; IN CORE INSTRUMENTS; NEUTRON FLUX; PHYSICAL RADIATION EFFECTS; RHENIUM ALLOYS; TEMPERATURE DEPENDENCE; THERMOCOUPLES; TUNGSTEN ALLOYS; VERY HIGH TEMPERATURE
Descriptors DEC
ALLOYS; MEASURING INSTRUMENTS; RADIATION EFFECTS; RADIATION FLUX; TRANSITION ELEMENT ALLOYS