Published February 1985 | Version v1
Journal article

Analytical study of performance of high resolution positron emission computer tomographs for animal study

  • 1. National Inst. of Radiological Sciences, Chiba-shi

Description

Performance of positron emission computed tomographs having spatial resolution as high as 3 mm FWHM was analytically evaluated. Increases in widths of line spread functions due to positron range and angular deviation were estimated. The slice thickness should be thin in accordance with the resolution in transverse axial plane. Singles rate, true coincidence rate, and scattered coincidence rate with high resolution tomographs of small diameter detector rings were estimated for three types of thin slice collimators. Feasibility of high resolution positron tomographs with small detector rings is discussed with applications to animal study

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-32
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
818-821
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Nuclear science symposium.
Dates
31 Oct - 2 Nov 1984.
Place
Orlando, FL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17044085
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ANIMALS; COLLIMATORS; COUNTING RATES; FEASIBILITY STUDIES; IMAGE PROCESSING; LINE BROADENING; LINE WIDTHS; PERFORMANCE; POSITRON COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; RESOLUTION; TOMOGRAPHY

Optional Information

Secondary number(s)
CONF-841007--.