4πβ-γ anti-coincidence absolute radioactivity measurement system

  • 1. China Inst. of Atomic Energy, Beijing (China)

Description

A 4πβ-γ anti-coincidence system has been developed for absolute radioactivity measurement. In this system, a dual ADC live-time HPGe γ-spectroscopy is used as the γ-ray detector, and a 4πβ proportional counter is employed for the β detector. The activity of 60Co, 134Cs and 166mHo have been measured. The combined standard uncertainty of 60Co, 134Cs and 166mHo, is 0.24%, 0.33%, 0.40%, respectively. The data of two former nuclei are compared with the results of 4πβ-γ coincidence counting system of NIM and CIAE and the agreement of these results is satisfactory. The decay-scheme of 166mHo is so complex that an adequate accuracy by using other methods is not acquired. It is also obtained that the result of anti-coincidence measurement is independent on β efficiency. At last, the relative questions in this study are also discussed