A glove box used at nuclear facilities must confine radioactive materials. High airtightness and negative internal pressure are used to prevent leaks. The allowable leakage rate of air is 0.1% vol/hr or less at the pre-service inspection. The negative pressure value is kept at - 30 mm H2O in normal operation. The glove box structural strength and its confinement reliability during an earthquake are major concerns. The verification of aseismic analysis methods and assumptions for a glove box are thus of great importance. Data on the dynamic behavior of giant glove boxes was recently obtained in large shaker experiments. This paper describes these experimental results and the appropriateness of aseismic analysis methods used in current design

Additional details

Publishing Information

Publisher
American Association for Structural Mechanics in Reactor Technology.
Imprint Place
Los Angeles, CA (USA)
ISBN
0-9623306-0-4
Imprint Title
Transactions of the 10th international conference on structural mechanics in reactor technology
Imprint Pagination
967 p.
Series
Seismic response analysis and design.
Journal Page Range
p. 877-886.

Conference

Title
10. international conference on Structural Mechanics in Reactor Technology (SMIRT).
Dates
14-18 Aug 1989.
Place
Anaheim, CA (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
21079688
Subject category
S42: ENGINEERING;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GLOVEBOXES; RADIATION PROTECTION; RADIOACTIVE MATERIALS; SEISMIC EFFECTS
Descriptors DEC
EQUIPMENT; LABORATORY EQUIPMENT; MATERIALS