Investigation of 3D diamond detector dosimetric characteristics

  • 1. Istituto Nazionale di Fisica Nucleare, Sezione di Perugia, Perugia (Italy)
  • 2. Dipartimento di Scienze Chirurgiche e Biomediche, Università degli studi di Perugia, Perugia (Italy)
  • 3. Dipartimento di Fisica Medica, Ospedale Santa Maria della Misericordia, Perugia (Italy)
  • 4. Istituto Nazionale di Fisica Nucleare, Sezione di Firenze, Firenze (Italy)

Description

Recently, a polycrystalline chemical vapor deposited (pCVD) 3D diamond detector with graphitic in bulk electrodes, fabricated using a pulsed laser technique has been evaluated for photon beam radiation dosimetry during in-air exposure. The same 3D diamond detector, has now been investigated to evaluate its performance under clinically relevant conditions putting the detector inside a Polymethylmethacrylate (PMMA) phantom, to obtain higher precision dosimetric measurements. The detector leakage current was of the order of ± 25 pA or less for bias voltages up to −100 V. The 3D detector was tested for time stability and repeatability showing excellent performance with less than 0.6% signal variation. It also showed a linear response for low dose rates with a deviation from linearity of 2%. It was also possible to verify the detector response as a function of the depth in PMMA up to 18 cm.