Published August 1996 | Version v1
Journal article

Evaluation of CdTe for use in a prototype emission/transmission CT imaging system

  • 1. Radiation Monitoring Devices, Inc., Watertown, MA (United States)
  • 2. Univ. of California, San Francisco, CA (United States). Dept. of Radiology

Description

Cadmium telluride has been investigated for potential use in a prototype imaging system capable of acquiring both x-ray CT and SPECT data. The system is being designed to accumulate SPECT images with 140 keV gamma rays, thus requiring 4 mm detector thicknesses to achieve satisfactory stopping power. This study primarily investigates whether an alternative pixel orientation can be used with thick detectors to preserve the higher photopeak efficiencies normally associated with smaller thicknesses. Using 57Co (122 keV) as a substitute isotope to 99mTc, small 2 mm CdTe cubes were investigated along with edge irradiated 2 x 4 x 2 mm bars of both CdTe and CdZnTe. The photopeak efficiency of the large bars can be increased through the use of cooling but energy resolution remains less than desired. Small cubes of either material meet the resolution requirements, but the efficiency of patient dose would be sacrificed

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
43
Journal Issue
4Pt1
Journal Page Range
p. 2225-2229.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
IEEE nuclear science symposium and medical imaging conference.
Dates
23-28 Oct 1995.
Place
San Francisco, CA (United States).

Optional Information

Secondary number(s)
CONF-951073--.