Elongation effects on the Therac 6 linear accelerator
Creators
- 1. London Clinic, Ontario Cancer Treatment and Research Foundation, London, Ontario, Canada and Department of Therapeutic Radiology, University of Western Ontario, London, Ontario, Canada
Description
The effects of field elongation on the central-axis output at d/sub max/ and on the depth-dose curves for the 6-MV x rays from the Therac 6 linear accelerator are considered. These parameters are independent of the collimator angle for both square and elongated fields. The exchange of collimator pairs results in significant output differences but has a negligible effect on the depth-dose curves. Both the central-axis output at d/sub max/ and the depth-dose curves agree satisfactorily with the predictions of the equivalent-square technique. An alternative method of greater accuracy is also indicated for the prediction of central-axis outputs at d/sub max/ for elongated fields