Published November 1986 | Version v1
Journal article

Dosimetry of asymmetric x-ray collimators

  • 1. Department of Therapeutic Radiology, University of Minnesota, Minneapolis, Minnesota 55455

Description

We have studied the dosimetry of an independent jaw system (provided with the Varian Clinac 2500) using ionometric measurements performed both in air and in a water phantom. Our study shows that the effect of the independent jaw on the dose distribution is similar to that of secondary blocking except for changes produced in the collimator scatter. A system of dose calculation was developed which takes into account the changes in the collimator scatter as well as in the isodose distribution. A method is described to correctly generate isodose curves for fields shaped by an independent jaw using a modified AECL TP11 treatment planning system. The primary modification in the program consists of correcting the zero-area tissue--maximum ratios for the off-axis variation in beam quality

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
13
Journal Issue
6
Series
Med. Phys.
Journal Page Range
936-941
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
18036531
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
AIR; ASYMMETRY; COLLIMATORS; ISODOSE CURVES; PHANTOMS; WATER; X-RAY DOSIMETRY; X-RAY EQUIPMENT
Descriptors DEC
DOSIMETRY; EQUIPMENT; FLUIDS; GASES; HYDROGEN COMPOUNDS; MOCKUP; OXYGEN COMPOUNDS; POLAR SOLVENTS; SOLVENTS; STRUCTURAL MODELS