Published September 2005 | Version v1
Journal article

Study on the internal quality control of TGAb and TMAb RIA in clinical laboratory

  • 1. RIA Center, Children's Hospital of First People's Hospital, Wuxi (China)

Description

To investigate the possible causes of the occasional instability of thyroglobulin anti- body(TGAb) and thyroid microsome antibody (TMAb) detection in the laboratory, the control serums of low, middle, and high value of different batches were measured and the stability of the labeled antigens stored at 2-8 degree C and below zero was observed. The results showed that the maximal variances of high value control serums for TGAb and TMAb between different batches were 33.0% and 32.5%, respectively. This was caused by the changes of titer of second anti- bodies. The labeled antigens stored below zero showed better stability. The strict internal quality control is an efficient measure to improve the measurement precision. (authors)

Additional details

Publishing Information

Journal Title
Labeled Immunoassays and Clinical Medicine
Journal Volume
12
Journal Issue
3
Journal Page Range
p. 158-160
ISSN
1006-1703

Optional Information

Notes
1 fig., 2 tabs., 1 ref.