Description
A permeameter for magnetically soft steels is described. The range of working induction values in the sample under investigation is from 1.6 to 1.95 T; working gap-50 mm; sample diameter-23 mm; sample length-230 mm. The measurement error of the relative magnetic permeability μr is ±1.5%. The block diagram of the permeameter, the schematic diagrams of its separate units and their brief description are presented
Availability note (English)
MF available from INIS under the Report Number.Files
25044512.pdf
Files
(331.5 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:06e1d4b4c18b95b1896323d705a01f99
|
331.5 kB | Preview Download |
System files
(36.4 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Пермеаметр для магнитомягких материалов
Publishing Information
- Imprint Pagination
- 14 p.
- Report number
- JINR-R--13-93-268
INIS
- Country of Publication
- Joint Institute for Nuclear Research (JINR)
- Country of Input or Organization
- Joint Institute for Nuclear Research (JINR)
- INIS RN
- 25044512
- Subject category
- S42: ENGINEERING;
- Descriptors DEI
- DIRECT CURRENT; ERRORS; FLOWSHEETS; INDUCTION; MAGNETIC SUSCEPTIBILITY; MEASURING INSTRUMENTS; SPECIFICATIONS; STEELS; TIMING CIRCUITS
- Descriptors DEC
- ALLOYS; CARBON ADDITIONS; CURRENTS; DIAGRAMS; ELECTRIC CURRENTS; ELECTRONIC CIRCUITS; INFORMATION; IRON ALLOYS; IRON BASE ALLOYS; MAGNETIC PROPERTIES; PHYSICAL PROPERTIES
Optional Information
- Notes
- 11 refs.; 7 figs.