Published January 1985 | Version v1
Journal article

Absolute neutron dosimetry: Effects of ionization chamber wall thickness

  • 1. Neutron Therapy Facility, Fermi National Accelerator Laboratory, Batavia, Illinois 60510

Description

To assess the effect of ionization chamber wall thickness on absolute neutron absorbed dose determinations, measurements were made of the charge collected by an A-150 tissue-equivalent plastic ionization chamber irradiated by a p(66)Be(49) neutron therapy beam as a function of chamber wall thickness both in air and in four different media: tissue-equivalent solution, water, motor oil, and glycerin. Wall thicknesses ranged from 1 to 31 mm, where isolation of the chamber gas volume from protons originating outside the chamber wall was assured. The in-air measurements compare favorably with earlier buildup measurements performed with an A-150 extrapolation chamber in an A-150 phantom. The in-phantom results may be explained if the effect of charged particles reaching the gas volume from the medium and the wall as well as the differences in neutron attenuation by the wall and the medium displaced by the wall are taken into account. The errors in absolute absorbed dose determination caused by ignoring the above processes are assessed

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
12
Journal Issue
1
Series
Med. Phys.
Journal Page Range
46-52
ISSN