Published April 1985 | Version v1
Journal article

An efficient algorithm for reconstruction of spect images in the presence of spatially varying attenuation

  • 1. Department of Nuclear Medicine, National Institutes of Health, ACRF, Bethesda, MD

Description

An algorithm is presented which permits the reconstruction of SPECT images in the presence of spatially varying attenuation. The algorithm considers the spatially variant attenuation as a perturbation of the constant attenuation case and computes a reconstructed image and a correction image to estimate the effects of this perturbation. The corrected image will be computed from these two images and is of comparable quality both visually and quantitatively to those simulated for zero or constant attenuation taken as standard reference images. In addition, the algorithm is time efficient, in that the time required is approximately 2.5 times that for a standard convolution-back projection algorithm

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-32
Journal Issue
2
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
1190-1197
ISSN
0018-9499
CODEN
IETNA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
18082540
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ALGORITHMS; ATTENUATION; IMAGE PROCESSING; SINGLE PHOTON ECT; SPATIAL RESOLUTION
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; RESOLUTION; TOMOGRAPHY