Published April 1990
| Version v1
Journal article
The HgI2 energy dispersive x-ray array detectors and minaturized processing electronics project
- 1. University of Southern California, Marina Del Rey, CA (USA). Inst. of Physics
- 2. X-Ray Instrumentation Associates, Menlo Park, CA (USA)
- 3. Stanford Univ., CA (USA). Stanford Synchrotron Radiation Lab.
Description
This paper describes recent progress in the development of HgI2 energy dispersive x-ray detector arrays for synchrotron radiation research and their associated miniaturized processing electronics. Deploying a 5 element HgI2 array detector under realistic operating conditions at SSRL, an energy resolution of 252 eV FWHM at 5.9 keV (Mn-Kα) was obtained. The authors also report energy resolution and throughput measurements versus input count rate. The results from the HgI2 system are then compared to those obtained under identical conditions from a commercial 13 element Ge detector array
Additional details
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 37
- Journal Issue
- 2
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 198-202
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- Institute for Electronic and Electrical Engineers (IEEE) nuclear science symposium.
- Dates
- 15-19 Jan 1990.
- Place
- San Francisco, CA (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 22004773
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPARATIVE EVALUATIONS; GE SEMICONDUCTOR DETECTORS; HGI2 SEMICONDUCTOR DETECTORS; MICROELECTRONICS; RESEARCH PROGRAMS; SYNCHROTRON RADIATION; X-RAY DETECTION
- Descriptors DEC
- BREMSSTRAHLUNG; DETECTION; ELECTROMAGNETIC RADIATION; EVALUATION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RADIATIONS; SEMICONDUCTOR DETECTORS
Optional Information
- Secondary number(s)
- CONF-900143--.