Published February 2010 | Version v1
Book

Boron-10 lined proportional counters for reactor instrumentation

  • 1. Electronics Division, Bhabha Atomic Research Centre, Mumbai (India)

Description

B-10 lined proportional counters for thermal neutron flux monitoring at the nuclear reactors have been developed. Design details of the counters with thermal neutron sensitivity of 20 cps/nv and 1 cps/nv are described. Systematic tests carried out on the detectors establish the usefulness for the reactor applications. The gas constants for the boron counter gas mixture 95% A + 5% CO2 have been evaluated. These values are extremely useful in estimation of the amplification factor for the boron lined counters at various operating parameters. The gamma tolerances of the detectors have been estimated experimentally. (author)

Part of:
Proceedings of DAE-BRNS national symposium on nuclear instrumentation - 2010

Additional details

Publishing Information

Publisher
Dept. of Atomic Energy
Imprint Place
Mumbai (India)
ISBN
81-88513-33-4
Imprint Title
Proceedings of DAE-BRNS national symposium on nuclear instrumentation - 2010
Imprint Pagination
679 p.
Journal Page Range
p. 302-306

Conference

Title
national symposium on nuclear instrumentation
Acronym
NSNI 2010
Dates
24-26 Feb 2010
Place
Mumbai (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
41082260
Subject category
S22: GENERAL STUDIES OF NUCLEAR REACTORS;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BORON LINED COUNTERS; NEUTRON FLUX; PERFORMANCE TESTING; REACTOR CORES; SPECIFICATIONS
Descriptors DEC
MEASURING INSTRUMENTS; NEUTRON DETECTORS; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RADIATION FLUX; REACTOR COMPONENTS; TESTING

Optional Information

Notes
9 figs., 2 tabs.